CO2 O2 UNIT BL CP
Supplies desired Air-CO2 or N2-O2-CO2 mixture at controlled pressure.

CO2-O2 UNIT BL CP
Features
- It replaces pre-mixed tanks
- One unit supplies up to 10 mini-incubators
- Mixes two or three gas, according to operator's need
Technical Specifications
- Operation mode: Adds CO2 and Air to N2
- Concentration Range: CO2: 4-8%; O2 3-7%
- CO2 Accuracy: ±0.1% after sensor calibration
- O2 Accuracy: ±0.1% after sensor calibration
- Set point resolution: 0.1%
- CO2 Repeatability: ±0.1%
- O2 Repeatability: ±0.1%
- Total Flow rate, Nl/min: 0.1-1.0 Nl/min (at 3.0 barg)
- Vessel Pressure: 3.0 barg
- Output Pressure (to mini-incubators): 0-3.0 barg
- Input Pressure: 4-4.5 barg
- CO2 Consumption Vessel 4.5L / Vessel 20L: 0.024 / 0.036 Nl/min at 6% CO2 and 0.3 Nl/min output flow rate
- N2 Consumption Vessel 4.5L / Vessel 20L: 0.281 / 0.421 Nl/min at 5% O2 and 0.3 Nl/min output flow rate
- Air Consumption Vessel 4.5L / Vessel 20L: 0.095 / 0.143 Nl/min at 5% O2 and 0.3 Nl/min output flow rate
- CO2 Sensor: Non Dispersive InfraRed (NDIR) dual wave length detector
- O2 Sensor: Zirconium oxide Sensors Life 10 years
- Input Gas connectors: 6 mm OD push in fittings
- Output Gas connectors: #4 Swagelok® ¼ in. OD connector - SS-400-1-2
- Controller Dimensions, mm: 440 x 410 x 207 (Width x Depth x Height)
- User interface: OKO-Touch
- Web Operation Through Smart Box
- Sensor filter: PTFE membrane with 0.2 μm pores
- Output gas filter: VOC and HEPA filter
- SDK Available for download, contact: info@oko-lab.com
- Sensor Calibration mechanism:
- ì) Comparison with external meter
- ìì) With Span Gas
- Pressurized vessel:
- ì) 4,5 l (CO2-O2 UNIT BL CP 4L). For up to 4 mini-incubators
- ìì) 20 l (CO2-O2 UNIT BL CP 20L). For more than 4 mini-incubators
- Recommended sensor calibration period: CO2 one year; O2 one year
- Certification: CE
[clone] CO2 O2 UNIT BL CP + SMART BOX + OKO TOUCH
[clone] Controller - Rear view