CO2 range: 0-20% - O2 range: 1-95%
![CO2-O2 UNIT-BL [0-20;1-95] CO2-O2 UNIT-BL [0-20;1-95]](/applications/webwork/site_okolab/local/document/001451.CO2-O2 Front_oko 006_thumb.jpg)
CO2-O2 UNIT-BL [0-20;1-95]
- Mixes CO2, O2 and N2
- CO2 range: 0-20%
- O2 range: 1-95%
- CO2 accuracy: ± 0.3 % at 5% CO2 and 5% O2
- O2 accuracy: ± 0.2% at 5% CO2 and 5% CO2
- Set Point Resolution: 0.1%
- Total flow rate: 0.35 Nl/min - constant
- Outlet pressure: ambient
- CO2 consumption (at 5% of CO2): 0.0175 Nl/min
- O2 consumption (at 5% of O2): 0.0175 Nl/min
- N2 consumption (at 5% of CO2 and at 5% of O2): 0.315 Nl/min
- Digital Flow Meters: CMOS Sensors
- Oko-Touch for intuitive touch screen operation
- Smart Box for data logging and remote operation
- Cabless connection to other units and to Smart Box
- Web operation through Smart Box
- SDK available for integration in third party software
- Can be attached to T Unit-BL H301 and T Unit BL-H201
CO2 O2 UNIT BL + T UNIT BL
In deep
CO2 O2 UNIT-BL CP
Supplies desired Air-CO2 or N2-O2-CO2 mixture at controlled pressure.